
Constant-power versus constant-voltage actuation in
frequency sweeps for acoustofluidic applications
Fabian Lickert,∗Henrik Bruus,†and Massimiliano Rossi‡
Department of Physics, Technical University of Denmark,
DTU Physics Building 309, DK-2800 Kongens Lyngby, Denmark
(Dated: 4 October 2022)
Supplying a piezoelectric transducer with constant voltage or constant power during a frequency
sweep can lead to different results in the determination of the acoustofluidic resonance frequencies,
which are observed when studying the acoustophoretic displacements and velocities of particles sus-
pended in a liquid-filled microchannel. In this work, three cases are considered: (1) Constant input
voltage into the power amplifier, (2) constant voltage across the piezoelectric transducer, and (3)
constant average power dissipation in the transducer. For each case, the measured and the simu-
lated responses are compared, and good agreement is obtained. It is shown that Case 1, the simplest
and most frequently used approach, is largely affected by the impedance of the used amplifier and
wiring, so it is therefore not suitable for a reproducible characterization of the intrinsic properties
of the acoustofluidic device. Case 2 strongly favors resonances at frequencies yielding the lowest
impedance of the piezoelectric transducer, so small details in the acoustic response at frequencies
far from the transducer resonance can easily be missed. Case 3 provides the most reliable approach,
revealing both the resonant frequency, where the power-efficiency is the highest, as well as other
secondary resonances across the spectrum.
Keywords: acoustofluidics, microparticle acoustophoresis, general defocusing particle tracking,
particle-velocity spectroscopy.
1. INTRODUCTION
In many experimental acoustofluidic platforms, the device is actuated by an attached
piezoelectric transducer, driven by a sine-wave generator through a power amplifier. To
describe the performance of the acoustofluidic actuation, the operating conditions are
typically expressed in terms of the voltage amplitude or the electric power dissipation
together with quantities such as the acoustic energy density, the acoustic focusing time,
or achievable flow rates [1–3]. Often, it is however left unclear under which conditions
and at which point in the electric circuit, the relevant quantities such as voltage am-
plitude or power dissipation have been measured. Recent studies compare device per-
formance at constant average power for different placements of the transducer [4,5].
Dubay et al. [6] performed thorough power and voltage measurements for the evaluation
of their acoustofluidic device, however, they noted that the actual power delivered to the
transducer might reduce to only a fraction (as low as 10%) of the reported value. The
likely cause of this reduction is that the transducer is acting as a large capacitive load,
where electrical impedance matching between source and load impedance is difficult to
accomplish [6,7].
Whereas optimization of the driving circuit is customary in other fields, such as ultra-
sonic transducers for cellular applications [8], non-destructive testing [9], and pulse-echo
systems [10], this has not been given much consideration in the field of acoustofluidics,
where the focus often lies on optimizing the acoustic impedance matching [11,12], while
neglecting the impact of the driving circuit. A recent work, though, considers topics
such as electrical impedance matching in the context of developing low-cost and possibly
hand-held driving circuits for acoustofluidics [13]. To our knowledge, studies have not
yet been performed, in which the impact of different electrical excitation methods on
∗fabianl@dtu.dk
†bruus@fysik.dtu.dk
‡rossi@fysik.dtu.dk
arXiv:2210.02311v1 [physics.flu-dyn] 5 Oct 2022