
tapered region connecting the conductors of the microstrip and CPW. The symmetric coupling on both sides of the wire
allows for differential readout. In combination with a differential amplifier, differential readout enables cancellation of
geometric jitter [
24
], common-mode noise rejection, and larger signal to noise, all of which can reduce overall timing
jitter.
The total inductance of the nanowire, which is dominated by the kinetic inductance of the superconducting NbN
(nanowire and waveguides), is a key design parameter that affects the maximum count rate and timing jitter. The
microstrips serve the purpose of increasing the kinetic inductance of the nanowires, to prevent latching. Short nanowires
are used in the design to minimize longitudinal geometric jitter [
7
], but the kinetic inductance of the
30 µm
-long nanowire
of
≈50
nH is too low, and would lead to latching of the device at low bias currents [
25
]. Each microstrip has a kinetic
inductance of
≈315
nH each, controlled by varying the width profile along the taper, with
Lk=Lsq R`
0
1
w(s)ds
, where
Lsq =194 pH is the sheet inductance, w(s)is the width along the microstrip (which varies between 120 nm and 1.5 µm),
and
`
is the arc length of the microstrip. The microstrips are meandered to allow them to be roughly equal in length,
leading to equal inductance across the 32 nanowires in the array. The sheet inductance is estimated from the reset time of
the device
τreset =Lk
RL
, which is measured in the Maximum Count Rate section. Device fabrication is described in the
Supplemental Materials.
Figure 1a also shows the overall shape of the detector chip. The circular area at the bottom in the picture is 2.5 mm in
diameter, designed to align with a standard FC/PC fiber ferrule. Self-alignment between the optical mode and the detector
active area [
26
] is enabled by a custom designed ceramic sleeve with a wider opening (1.9 mm) that fits around both
ferrule and detector.
Figure 1b shows the normalized photon count rate (PCR) versus bias current for nanowire 16. This measurement is
done at low count rates – here 82 kcps at saturation. The saturation of the count rate for currents above
10.5 µA
indicates
unity internal detection efficiency (IDE), which means that any photon absorbed in the nanowire will create an output
pulse that can be detected [
27
]. The PCR curve was fit to an idealized model of the IDE as a function of current in the
nanowire,
η=1
2erfc−Inanowire−Idetect
σ
, where
Idetect
is the inflection point and
σ
is a fit parameter describing the slope of
the inflection. At low count rates the current in the nanowire is assumed to be equal to the bias current (
Inanowire =Ibias
)
meaning the nanowire is not recovering from a previous detection when it absorbs a photon. Figure 1b also shows the dark
count rate (DCR) of the same nanowire. In this measurement, dark counts are mostly due to ambient photons coupling to
the detector, as indicated by the plateau of 0.4 counts/s at bias currents below
Iswitch
. No filtering or special enclosure was
used to minimize the dark count rate. Shortly before
Iswitch
, the DCR shows the typical exponential increase. After
Iswitch
the nanowire enters a regime of relaxation oscillations, then the count rate starts decreasing, indicating latching. We define
Ilatch
as the nanowire current at which DCR starts to decrease. Figure 1c shows
Idetect
and
Iswitch
for all nanowires.
Idetect
values were similar, indicating that the nanowires are fairly uniform across the array, though the slightly higher variation
in switching current indicates varying degree of constriction in the wires. All wires had large plateaus in IDE vs. current.
Since the IDE exceeds
90%
for
Ibias >Idetect +σ
, a measure of the plateau is
Iswitch −(Idetect +σ)
, which ranged from
0.7 µA
to
3.9 µA
for the wires in the array, with an average of
2.8µA
. In order to take advantage of this long detection
plateau, the nanowires were biased at 95% of their individual switching currents for all measurements.
3 MEASUREMENT SETUP
The PEACOQ array was measured in two different cryostats at a temperature of
0.9
K. One cryostat (Setup A), was
used to read out all 32 channels simultaneously. Setup A was initially developed as a testbed for the Deep Space Optical
Communications project [
28
]. The second cryostat (Setup B) was used to perform low-jitter measurements of a single
detector channel. Further details of this setup can be found in
7
. In each case, the device was wire-bonded to a custom
PCB board, designed for differential read-out of each of the 32 nanowires, and packaged to accommodate self-aligned
fiber coupling [
26
]. One side of each nanowire was terminated with a
50 Ω
resistor at
0.9
K for a single-ended readout of
each wire. Future work will focus on a fully differential readout.
In Setup A, microcoaxial cables connected the signal from each nanowire to a cryogenic readout board at 40 K. The
cryogenic readout board contained a resistive bias-tee and a two-stage cryogenic amplifier (DC-coupled HEMT and
SiGe LNA) for each channel. At room temperature, the pulses were further amplified by a third stage of amplifiers to an
amplitude of
≈500
mV, then converted to time stamps using a custom 128-channel time-to-digital converter (TDC) from
Dotfast Consulting. The optical source comprised either a continuous or pulsed laser at 1550 nm, a variable attenuator,
polarization controller, and a switch that sent light either to the PEACOQ or to a power meter. An SMF-28 optical fiber
was used to couple light to the detector. The face of the optical fiber ferrule had an anti-reflective coating optimized for
1550 nm. The pulsed laser had a pulse width of
0.5
ps and a repetition rate of 20 MHz. Using a phase-locked loop, the
sync signal of the pulsed laser was converted to a 10 MHz signal to synchronize the TDC. The timing jitter of the readout
3/12